To view detailed settings for a Transfer Function measurement engine, double-click its control block on the Control Bar or open the Measurement Config panel and either double-click its color tile in the Measurement Configuration table or select its name and click the Info button below the table. Any of these actions will take you to a page with the settings below.
.png?sv=2026-02-06&spr=https&st=2026-11-18T17%3A30%3A01Z&se=2026-11-18T17%3A48%3A01Z&sr=b&sp=r&sig=4w%2B%2FTi6%2BQKeyM%2B081jlTRbmjFmXieCeLQnuCvrTDfQY%3D)
Transfer Function Measurement Engine Settings flyup menu.
Measurement Settings
The Name field under Measurement Settings sets the measurement name. If you edit the measurement name, be sure to press the [Enter] key to set the change when you are finished.
The Delay field sets the amount of Signal delay (in milliseconds) needed to align the reference and measurement signals. Positive values delay the measurement signal (the most common case). Entering a negative value delays the reference signal.
FFT size (in samples) determines the Time Constant and frequency resolution of the frequency Domain displays. When MTW is selected, Smaart performs multiple FFTs and combines the results into a single frequency data set. The individual FFT sizes used in this case are not user selectable. (Suite and RT only)
Averaging Depth specifies the length of time over which individual measurements are averaged to stabilize transfer function traces on the analyzer screen and improve their signal-to-noise ratio. Bigger numbers give you longer averaging times.
Avgerage Type sets the type of averaging used for Magnitude traces. The options are Polar (RMS) or complex (vector). Phase traces always use complex averaging. Polar averaging is probably the most common type for magnitude traces, but both types have their uses. In practical terms, Polar averaging lets more reverberant energy into the average, which may tend to agree better with what you hear, particularly for musical program material. Complex magnitude averaging tends to reject reverberant energy as noise and may give you better clues regarding speech intelligibility than polar averaging.
The FFT, Averaging Depth, and Average Type settings follow the global settings set in the Transfer Function Measurement Settings dialog. You can however, set any of these for an individual measurement engine by deselecting its Use Global checkbox and selecting a new setting from the drop-list.
Display Settings
Color has a clickable tile that, when clicked, opens a color picker dialog to change the display color for the measurement. When committed, these changes apply to all linked copies of the measurement with the same name on other tabs and windows.
Plot sets the preferred graph assignment for the measurement. It is ignored when only one graph of a given type is displayed. If you bring up a second graph of the same type in the same tab, measurements with a Plot setting of 1 (the default setting) will stay with the first instance of the graph and any measurements with a setting of 2 will move to the second.
Invert Magnitude Display, when checked, will cause Smaart to plot the measurement’s magnitude response upside-down. This can be a handy option when setting loudspeaker EQ curves – a cut filter will plot as a boost and a boost filter as a cut when the measurement is inverted, making it easy to create a complimentary EQ curve.
The following settings can be assigned locally or at the graph level for all displayed live measurements. If the Follow Plot check box is checked for any of the following parameters, they will follow changes to the settings in the Transfer Function Graph Settings dialog and/or on-graph controls. If not, the measurement will keep its own settings and ignore the graph-level settings.
Phase Smoothing sets the amount of smoothing used for the transfer function Phase display. This option can be set globally or locally for each measurement.
Mag Smoothing sets the degree of smoothing type for transfer function magnitude traces. This option can be set globally or locally for each measurement.
Weighting applies a weighting curve to the measurement to reshape its response curve, subtracting from the magnitude values at some frequencies and perhaps adding to them at others. Weighting in the frequency domain is analogous to filtering the measurement signal in the time domain. Common weighting curves include A and C weighting used for SPL and Leq measurements. (Suite and RT only)
Input Settings
This section allows you to make changes to the measurement engine’s routing.
Device - The I-O device (or driver).
Channel - The specific input from said I-O device that is being used.
Measurement Signal - Typically the input corresponding to your measurement mic, used to capture output data from the System Under Test.
Reference Signal - The signal against which your Measurement Signal is compared. This is typically an input receiving your I-O device’s output signal. This can either be accomplished via a physical loopback cable connecting one of your I-O device’s output channels into one of its input channels, or via a digital loopback originating from the I-O device itself or some sort of routing application. Smaart can also facilitate a digital loopback, if desired.
Normally, both Measurement and Reference signals will come from the same input device, so the Device selection for the Measurement Signal is automatically applied to the Reference Signal as well.
Multi-Device TF:
It is possible to use signals from two different devices if you enable Allow Multi-Device TF in Options > Preferences > Advanced, but this will only work if both devices’ sample clocks are synchronized. If the devices are out of sync, you will see your measurement’s delay time drift. Additionally, it is important to note that different devices will apply different processing delays to each signal path. As such, a Multi-Device TF will often calculate delay times that don’t necessarily correlate to physical distance.
Global TF Settings
Two of the settings in the Global Spectrum Settings control group, Mag Threshold and Blanking Threshold, apply to all transfer function measurements in your configuration. The rest can be all applied globally or localized for individual measurements by un-checking their Use Global check boxes in the Measurement Settings control group.
Mag Threshold sets a normalized Full Scale (dBFS) value that the reference signal must exceed before new data is accepted into the measurement at any given frequency. When the magnitude of the reference signal does not cross threshold at some frequency, new incoming data at that frequency is excluded from the average.
Blanking Threshold sets the coherence value that must be met or exceeded before a data point at a given frequency is displayed on the graph. This setting applies to both phase and magnitude graphs.
The following settings apply to any transfer function measurement that subscribes to the global settings for that parameter: FFT, Averaging, Mag Avg Type, Phase Smoothing, Mag Smoothing, and Weighting.