What is an SPL Calibrator?

Prev Next

An acoustic calibrator is a specialized piece of hardware used to calibrate Smaart for SPL. It provides a test tone at a specified level. Once used during Smaart’s calibration procedure, the software will produce SPL readings relative to the level of this test tone.

It operates rather simply:

  1. Place your measurement microphone capsule snugly inside of the port on the side of the calibrator. Be sure to use the correct size adapter for your microphone. A snug fit is required to create a seal. If this seal isn’t created, it can lead to inaccurate calibration. Many calibrators come with one or more adapters, and some microphone manufacturers offer microphone-specific calibrators. We offer a selection of  adapters through the Rational Acoustics online store.

  2. Once your microphone is in the calibrator’s port, turn on the calibrator.

  3. When turned on, the calibrator emits a 1 kHz sine wave reference tone at a specified level (typically 94 dB, but sometimes 114 dB or 110 dB). There are some calibrators that have the ability to switch between two or more levels. With these models, it is paramount to ensure the proper level is selected so as to not skew your readings.

  4. Complete the calibration steps in Smaart to complete this process. Smaart will read the dBFS level of the calibrator’s reference tone coming through the microphone and associate it to the level you’ve input in the “Calibrator Level” field in Smaart. By default it is set to 94dB, which is the most common calibrator level. Once calibrated, Smaart interprets all SPL readings relative to this reference level.

While there are many calibrators available, we'll be outlining the 3 models of calibrator we have on the Rational Acoustics online store: The iSEMcon SC-1, the Cirrus CR:514 (Class 2), and the Cirrus CR:515 (Class 1). For more information on the differences between Class 1 and Class 2 measurement systems, click here.


iSEMcon SC-1

The iSEMcon SC-1 is a single-frequency, dual-Amplitude calibrator that generates a 1 kHz reference tone at either 94 or 110 dB SPL. It ships with both a 1/2" and a 1/4" mic adapter and requires a 9 Volt battery for power.


Cirrus CR:514 and CR:515

The Cirrus CR: 515 is designed specifically for Class 1 measurement devices in order comply with the current IEC 60942:2017 standard for acoustic calibrators with Class 1 specifications. It has a 1/2" port and ships with a 1/4" adapter. It emits a 1 kHz sine wave reference Signal at a fixed level of 94 dB SPL.

The Cirrus CR:514 also produces a 1kHz reference signal at 94 dB SPL, but is engineered for current IEC 60942:2017 Class 2 measurement standards. Like the CR: 515, its port is 1/2" and it ships with a 1/4" adapter.