To calibrate one or more input channels for sound level (SPL or Leq) measurement in Smaart, press [Alt] + [A] on your keyboard or select I-O Config from the Config menu to open the Configurator dialog to the I-O Config page. In I-O Config, select the input device and channel that you want to calibrate and click the Calibrate button at the bottom of the page below the channels table. This opens the Amplitude Calibration dialog. At the top of the calibration dialog, make sure that the device and channel that you want to calibrate are selected on the Input Device and Input Channel drop-list selectors.
Calibrating with a Sound Level Calibrator
In the realm of digital audio, Amplitude values are abstracted from absolute references such as volts or pascals. Sample values tell us nothing about the maximum voltage swing of the A/D converter, microphone or preamp sensitivity or anything else about what was connected to the A/D's inputs, and so the common convention is to take the biggest number we can get from a digitized signal (e.g., given a sample word of a given number of bits in the case of a PCM signal) and scale that number to ± 1.0 in peak linear amplitude terms. In Decibel terms, that means 0 dB FS will be the maximum possible Magnitude and all lesser magnitudes are negative numbers.
To relate these internal full-scale amplitudes to some number of volts or Pascals of pressure in the real world, we need to reference to a signal of known physical amplitude – and to do that, when there is a microphone in the signal path, we need an acoustical signal of known amplitude. This is where a sound level calibrator comes in.
Calibrating an input channel and microphone for SPL measurement using a sound level calibrator is a two-step process consisting of:
Measuring the digital Full Scale signal level of a microphone with the sound level calibrator coupled to it, and
Assigning the reference sound level for the calibrator (typically 94 or 114 dB) to the measured Full Scale amplitude in Smaart
To measure the input signal level, you need to:
Connect a microphone to the input channel that you want to calibrate.
Affix your sound level calibrator to the microphone.
Turn on the calibrator and adjust the gain for your input channel to a desirable level.
Set your Calibrator Level in the Amplitude Calibration dialog (if different from Smaart's default, 94 dB).
Click the Calibrate button in the Amplitude Calibration dialog to run Smaart’s calibration routine.
Bear in mind, however, that once you are calibrated, you will need to keep the input gain set exactly where it is to maintain calibration, and so a little bit of forethought with regard to setting levels might save you needing to repeat this procedure again later.
Before calibrating for SPL, consider the highest level you will need to measure, and make sure your mic has a sufficient Max SPL rating. For example, for measuring SPL at concert levels, we recommend that your microphone can accommodate a Max SPL of at least 135 to 140 dB SPL.
It is also important that your gain setting is low enough to accommodate your target SPL without overloading. The Amplitude Calibration dialog has an input level meter that shows you the peak full-scale signal level for the selected channel). With your microphone calibrator running, adjust the gain for the input channel to your target full-scale amplitude level. For example, if your SPL target is 140, and your calibrator produces a 114 dB SPL reference tone, adjust the gain to leave at least 26 dB of headroom above the calibrator reference level.
When your levels are set, click the Calibrate button. The Calibration Progress dialog pops up, Smaart measures the input signal over a period of a few seconds, and then reports the full-scale signal level. If you are happy with the result, then click OK.
Back in the Amplitude Calibration dialog, you will see that Smaart has calculated the necessary Offset value in dB to calibrate the selected input to SPL. The input and microphone are now calibrated for sound level measurement. You can click OK to exit the dialog or select another input to calibrate.
Note that it is possible to simply type a calibration offset value directly into the Offset dialog without running the Calibrate routine if you know the offset value and you are certain neither the microphone sensitivity nor the gain of the input channel have changed since a previous calibration. In this case, be sure to press the [Enter] key on your keyboard after typing in the value to apply the change.
Gain Tracking
When the selected input device is one of our supported integrations, Smaart will detect changes and automatically adjust the calibration settings for calibrated input channels to compensate, provided that the Gain Tracking check box for the device, located under the input channels table for the device in I-O Config, is checked. The accuracy of these adjustments is device-dependent but will typically range within a maximum error of about ± 1.5 dB from nominal gain setting, between 0 and 50 dB, when calibrated at a gain setting of 25 dB.
To quickly determine the maximum gain adjustment error for your specific device, set input to a gain to 25 dB, then calibrate for sound level measurements using a sound level calibrator (see above) with Gain Tracking turned off. After calibrating, with the calibrator still fitted to the microphone and turned on, enable Gain Tracking and assign the input to an SPL meter in Smaart, then observe the measured level as you turn the gain up to 50, and down to 0 dB. The maximum deviation observed from the reference level of the calibrator represents the maximum expected accuracy of your setup.
Supported Devices for Gain Tracking
Smaart v9 supports the following devices for gain and preamp control, as well as gain tracking:
As of Smaart v9.6 there are three types of I-O Device that support direct Smaart integration:
Audient EVO 4, 8, & 16 (v9.1 or later. Audient EVO SP8 is not supported)
Roland Octa-capture (Smaart v8 or later. These devices are discontinued)